ACEY-HRCDS-6V500A-4CH battery life cycle tester supports battery cycle life test, battery capacity test, battery charge discharge characteristic test, battery charge retention ability test, battery charge and discharge efficiency test, battery overcharge and over-discharge rate tolerance test, pulse test, operating condition test, etc.
Brand:
ACEYItem No.:
ACEY-HRCDS-6V500A-4CHCompliance:
CE CertifiedWarranty:
One Year warranty with lifetime supportOrder(Moq):
1 SetPayment:
T/TProduct Origin:
ChinaLead Time:
7days6V 500A Regenerative Lithium Battery Life Cycle Tester
Product Description
ACEY-HRCDS-6V500A-4CH Lithium battery life cycle tester is mainly used in life aging test and quality control in battery production or experiment including lithium-ion batteries, lead-acid batteries, nickel-cadmium batteries, nickel-metal hydride batteries, etc. Energy feedback type, bidirectional energy saving of charge and discharge, saving a lot of power consumption expenses.
Product Features
Working Environment and Technical Parameters
Parameters
No.
Item
ACEY-HRCDS-6V500A-4CH
1
Main channels
2CH/ module
2
Channel working mode
Completely independent.
3
Main channel working mode
Charge: CC, CV, CCCV, no voltage or current impact when CC transfer to CV.
Discharge: CD, CP, CR, DCCV.
Still, Pulse test, Working Condition test, DCIR test.
Main channel test cutoff condition
Time, total voltage, total current, capacity, -ΔV, Vcell, RSOC, etc.
4
Main channel Voltage
Single channel test range
-6V~6V
Min. discharge voltage
-6V (device port voltage)
Accuracy
≤0.05%FS(@25℃±10℃),≤0.1%FS(@0-15℃,35-45℃)
Resolution
1mV
Main channel Current
Single channel test range
±500A
Min. output current
50mA
Accuracy
50mA--50A:0.05%FS(@25℃±10℃)
50A--100A:0.05%FS(@25℃±10℃)
100A--500A:0.05%FS(@25℃±10℃)
Resolution
1mA
Current start response time
<5ms(10%FS~90%FS)
Time
Resolution
10mS
Auxiliary temperature (Optional)
Range
-20℃~100℃
Accuracy
±1℃(0℃~70℃)
Resolution
0.1℃
Power
Output power
Single channel continuous working power 2500W
Stability
±0.1%FS
5
Communication
LAN
6
Control program
Support data protection when power off, continue testing caused of power cutoff or manually stop and revise step parameter online.
Users can set protection condition, include minimum voltage, maximum voltage, minimum current, maximum current, etc.
Users can set auxiliary channel protection condition, including minimum voltage, maximum voltage, maximum temperature, minimum temperature.
Cycle time
Max. 9999
Programmable step
Max. 9999
Nested loop
Max. 10 layers
Step time range
No limit, support h, min, s
Data record
Maximum recording time: 10ms, voltage ≤0.05%FS, current ≤0.05%FS
Program feature
Each step has one or multiple exit
7
Data display
Cycle list
Circulation number. Charge/discharge capacity, efficiency, power and average voltage, etc.
Process list
Process number, working mode, process time, capacity, power, DIR, medium voltage, cut-off voltage, cut-off current, etc.
Detail list
Record number, system time, cumulative time, voltage, current, energy, power, cell temperature, etc.
8
Data export
Data can be exported as EXCEL, TXT or diagram. Provide charge discharge efficiency, charge discharge curve, voltage, capacity, current-time characteristic curves, etc.
9
Data saving
Save as files
10
Curves type
X-coordinate: total time, charge discharge capacity, cycle time, voltage, current, capacity.
Y-coordinate: total voltage, current, capacity, power, cell temperature(Optional), end of charge and discharge capacity, etc.
11
Software protection
Power off data protection, power resume continue testing, OVP, ODP, OCP, UCP, OTP, capacity protection, abnormal voltage trend protection, abnormal current trend protection, abnormal current fluctuation protection, abnormal voltage fluctuation protection, abnormal charging voltage increase rate protection, abnormal discharge current decrease rate protection.
Hardware protection
Power off, open circuit and short circuit protection. Reverse or wrong connection protection, OCP, OVP, OTP, fixture contact abnormal protection.
Warning function
Switch off automatically under emergency, shutdown when power off, load connect and disconnect automatically.
12
Barcode scan
Support barcode start and binding test data, classify batteries rapidly and accurately. Support detailed battery database management functions, connect battery information with testing data base, which is easy for management of data base system.
13
Sampling terminals
Four-polar sampling. Copper nose or other fixture.
14
Energy consumption
Discharge power feedback to grid
15
Cooling method
Forced Fan-cooled
Noise
Less than 70dB
IP grade
IP20
16
Repair
Modularized structure. Can be replaced or repair with spare parts.
17
Software update
Provide regular software upgrade services until the hardware is incompatible.
Equipment Specification
Cabinet size | 600*850*1030mm |
Weight | ≤250KG |
Channels number | 4CH |
Remarks :The above picture is for reference only, subject to the kind! |
Product Display